Abstract
A coated microneedle array comprises sharp micrometer-sized needle shafts attached to a base substrate and coated with a drug on their surfaces. Coated microneedles are under investigation for drug delivery into the skin and other tissues, and a broad assortment of active materials, including small molecules, peptides, proteins, deoxyribonucleic acids, and viruses, have been coated onto microneedles. To coat the microneedles, different methods have been developed. Some coating methods achieve selective coating of just the microneedle shafts, whereas other methods coat not only microneedle shafts but also the array base substrate. Selective coating of just the microneedle shafts is more desirable since it provides control over drug dosage, prevents drug waste, and offers high delivery efficiency. Different excipients are added to the coating liquid to modulate its viscosity and surface tension in order to achieve uniform coatings on microneedles. Coated microneedles have been used in a broad range of biomedical applications. To highlight these different applications, a table summarizing the different active materials and the amounts coated on microneedles is provided. We also discuss factors that should be considered when deciding suitability of coated microneedles for new-drug delivery applications. In recent years, many coated microneedles have been investigated in human clinical trials, and there is now a strong effort to bring the first coated microneedle-based product to market.
Introduction
Microneedles are sharp protrusions measuring less than a millimeter in length. Microneedles were originally invented for drug and vaccine delivery through the skin (Henry et al., 1998). Microneedles achieve this objective by piercing the top layer of the skin, the stratum corneum, which is the principal barrier to transdermal delivery of drugs (Bouwstra and Ponec, 2006). Microneedles are attractive because they are minimally painful (Gill et al., 2008; Haq et al., 2009; Nguyen and Park, 2018) and can be applied by patients in the comfort of their homes (Birchall et al., 2011). A microneedle array (also known as a microneedle patch) comprises one or more individual microneedles situated on a flat base (also called a substrate). When the individual microneedles rise perpendiclular to the flat base, it is called a two-dimensional array, and when the individual microneedle run parallel to the flat base mimicking a “fork”, it is called a one-dimensional array. On the basis of their design and mode of use, microneedles can be categorized into four basic types: hollow microneedles, dissolvable microneedles, poke-and-patch microneedles, and coated microneedles. These classes have been reviewed before (Prausnitz, 2004; Quinn et al., 2014), and we only briefly introduce them in this work. A hollow microneedle is similar to a conventional hypodermic needle but is notably much shorter. It has a hollow bore and liquid can be injected through it. A dissolvable microneedle on the other hand is made entirely of either drug or of a matrix in which the drug is dispersed. The dissolvable microneedle upon insertion into the skin can dissolve within minutes or can degrade over a longer time. In the “poke-and-patch” approach, a solid microneedle is used to pretreat the skin to increase its permeability by creating a micrometer-sized perforation, after which the drug formulation or a transdermal patch i applied over the pretreated skin. Finally, with a coated microneedle, the drug is coated as a solid film on the microneedle surface. When the coated microneedle is inserted into the skin it carries the coated payload with it. Once in the skin, within minutes the coating dissolves, after which the microneedles can be removed and safely discarded.
A coated microneedle is a versatile delivery system. The same microneedle patch can be used for the delivery of a broad spectrum of materials ranging from small molecules, to proteins, to DNA, to viruses, and even to microparticles (Gill and Prausnitz, 2007a). In contrast to the dissolvable microneedle, whose mechanical properties can change when the encapsulated drug fraction is altered or a different drug is dispersed in its matrix, the mechanical properties of a solid microneedle are not impacted when a different drug is coated on it surface. Furthermore, the coated microneedle system has been evaluated as a delivery system for not just the skin (Prausnitz, 2017) but also the eye (Jiang et al., 2007; Kim et al., 2014; Song et al., 2015), vascular tissue (Lee et al., 2014, 2017b), and the oral cavity (McNeilly et al., 2014; Ma et al., 2015). A recent article by Rzhevskiy et al. (2018) has reviewed the use of microneedles for drug delivery into different tissues and organs.
To use coated microneedles for different applications, it is important to consider the quality of coatings, reproducibility of the coating process, and the delivery efficiency. In the first publication on coated microneedles (Matriano et al., 2002), it was reported that the microneedle patches were coated by immersion in a coating solution. This resulted in coating of the entire microneedle array, including the base substrate. A few years later, to improve delivery efficiencies and to reduce drug wastage, more precise coating methods were developed to restrict coatings to just the microneedle shafts; one such method was published in 2004 (Cormier et al., 2004) and another in 2007 (Gill and Prausnitz, 2007a). Early studies also described the influence of liquid properties on the uniformity of the coating of microneedles and demonstrated that a broad range of materials could be coated onto microneedles with precision (Gill and Prausnitz, 2007b). Since the publication of these early studies, many microneedle coating methods have been developed. The objective of our review is to highlight the important design criteria and parameters that should be considered in the development of coated microneedles and to review the different microneedle coating methods. A comprehensive analysis of in vivo and in vitro results from the application of coated microneedles is outside the scope of this review. We refer the readers to the recent microneedle review articles for this information (Prausnitz, 2017; Shin et al., 2017; Nguyen and Park, 2018; Rodgers et al., 2018).
The Concept of Coated Microneedles
A coated microneedle comprises a sharp, solid-core microneedle structure on which is coated a solid film containing the active compound and water-soluble inactive excipients. The water-soluble excipients not only aid in the microneedle coating process but also catalyze the detachment of the film from the microneedle surface. The process by which a coated microneedle delivers its payload into the skin is schematically shown in Fig. 1. When a drug-coated microneedle is inserted into the skin, the coating encounters the interstitial fluid present in the tissue. Contact with this aqueous medium helps to dissolve the water-soluble excipients in the microneedle coating, which in turn initiates the detachment of the coating from the microneedle surface. Depending on the aqueous solubility of the coating excipients, the detachment of the coating can finish within seconds to minutes. It is only necessary that the coating detach from the microneedle surface before microneedles are removed from the skin; the complete dissolution of the material left behind can occur in due time. Coated microneedles can be used to deliver their coated payload not only to skin but also other tissues.
Desirable Attributes in Microneedle Coatings
Uniform Coatings.
To ensure reproducible delivery of active molecules, it is important to produce uniform coatings on microneedle surfaces. If as depicted in Fig. 2-i, coatings are irregular, random, and patchy in nature, poor consistency in the mass of material being coated is achieved. Therefore, a primary attribute required in microneedle coatings is that they should form a uniform coverage on the microneedle shafts without uncoated spots.
No Deposition of Coatings on the Base Substrate of the Microneedle Patch.
Only microneedle shafts penetrate the skin. As such, drug that is not coated on the microneedle shafts but is instead situated on the base substrate of the microneedle patch (Fig. 2-ii and 2-iii) is not deliverable into the skin. This wasted coating located between the microneedles can be quite significant. For example, a 10 × 10 mm patch with 10,000 microneedles/cm2, each measuring 250 μm in length and 40 μm in base diameter and tapering into a sharp point of less than 2 μm (Griffin et al., 2017; Fernando et al., 2018) has a total surface area of about 245 mm2, of which about 65% (158 mm2, assuming conical geometry) is useful since it is located on microneedle surfaces, whereas 87 mm2 (35%) constitutes the wasted coating on the base substrate. The most preferred scenario for producing coatings on microneedles is shown in Fig. 2-iv, wherein the coatings are not only uniform but are also situated only on the microneedle shafts. This can allow for a reproducible and a highly efficient coated microneedle delivery system.
Coating Methods
To coat microneedle surfaces with drug molecules, the two must be brought in physical contact with each other. Depending on how this contact is achieved, the coating methods can be classified into dip coating, inkjet coating, immersion coating, drop coating, and spray coating. These methods vary greatly in their ability to selectively coat the microneedles and can be further categorized into two groups: The first group of methods can generate coatings selectively on the microneedle shafts without coating the base substrate of the microneedle array, and the second group of coating methods result in coating of both the microneedle shafts and the base substrate of the microneedle array.
Coating Methods That Selectively Coat the Microneedle Shafts without Contaminating the Base Substrate of the Microneedle Array
Dip Coating.
In this method, a microneedle is dipped into the coating liquid. The microneedle, upon exiting the coating liquid entrains a liquid film on its surface. When the solvent in the liquid film evaporates, the dissolved or suspended solids that are present in the liquid film get deposited as a coating on the microneedle surface. However, at the micrometer length scale of the microneedles, contribution of the surface tension of the coating liquid becomes significant. Owing to surface tension–induced capillary action, the coating liquid rises along the microneedle length and the meniscus can touch the flat substrate of the microneedle shafts leading to contamination of the base substrate with the coating liquid. As discussed in the previous section, this is an undesirable outcome. To overcome this issue, two general approaches have been developed to selectively coat just the microneedle shafts. In the first approach a masking plate is used to prevent the base substrate from touching the coating liquid and to allow the microneedles to contact only the coating liquid. In the second approach a thin film of the coating liquid is created into which the microneedles are dipped. In particular, the thickness (depth) of the coating liquid is kept lower than the height of the microneedles. As a result, when the microneedles are dipped into the thin liquid film, the coating liquid is unable to rise significantly and thus is unable to touch the base substrate of the microneedle array. These two techniques are elaborated below.
Masked dip coating.
A physical mask can be introduced between the microneedle base substrate and the coating liquid interface. The mask is designed to allow contact of microneedles with the coating liquid but to physically keep the flat substrate of the microneedle patch from touching the liquid. The mask concept was first used by Gill and Prausnitz (2007a) (Fig. 3A-i), wherein the dip-holes in the mask were produced via laser machining, and later produced using microfabrication techniques by Caudill et al. (2018). Gill et al. laser machined a channel to hold the coating liquid and covered it with a masking plate with laser-drilled micrometer-sized holes (Fig. 3A-ii). The holes were made to precisely correspond to the locations of the microneedles in the array. After proper alignment, the microneedle arrays were successfully dip-coated without contaminating the base substrate. Figure 3A-ii shows a one-dimensional array being dip-coated, and Fig. 3A-iii shows a two-dimensional microneedle array coated with this approach. Instead of laser machining, Caudill et al., used the continuous additive manufacturing technique called continuous liquid interface production (CLIP) to fabricate both the microneedles and the masked dip-coating system (Caudill et al., 2018) (Fig. 3B-i), and successfully coated just the microneedle shafts (Fig. 3B-ii). An advantage of the masked dip-coating process is the fine control over the length of the microneedle that can be coated. Both Gill and Prausnitz (2007a) and Caudill et al. (2018) demonstrated this fine control by coating from as little as 25% to almost 100% of the microneedle shaft without contaminating the base substrate of the microneedle patch. The dip-coating technique with a mask has been used by other investigators (Andrianov et al., 2009). In a variation of the masked dip-coating method, a single orifice, which is often a small-diameter pipette tip (Chong et al., 2013) or a hypodermic needle (Witting et al., 2015) has also been used to dip coat the microneedles. Conceptually, the single orifice of the pipette or the hypodermic needle can be considered equivalent to a single masked hole that allows selective contact of the microneedle with the coating solution.
Thin-film dip coating.
Besides using a mask, capillary rise can also be countered by reducing the thickness (depth) of the coating liquid. Dip coating of microneedles into a thin film of the coating liquid whose thickness measures less than the height of the microneedle shaft diminishes the capillary rise. Two main approaches have been used to generate the thin liquid film. In one approach a drum is partially submerged in the coating liquid and rotated (Fig. 3C-i). As the drum rotates a thin film of liquid is entrained on the drum circumference and its thickness is controlled with the help of a doctor blade. Microneedles are moved tangential to this liquid film and dipped into the film to achieve coatings (Cormier et al., 2004; Ameri et al., 2010). Ameri et al. placed a roller drum in a drug formulation reservoir (2 ml in volume) and rotated the roller at 50 rpm to generate a thin film of the coating liquid (approx. 100 μm) and dipped 100-μm long microneedles into this thin film to selectively coat just the microneedle shafts (Fig. 3C-ii) (Ameri et al., 2010). In the second approach, a thin film is prepared by using a shallow reservoir whose depth is less than the length of the microneedle shaft, and microneedles are subsequently dipped into this liquid film for coating (Fig. 3D-i). Different studies have used this coating approach (Zhang et al., 2012a,b; Kommareddy et al., 2013; Chen et al., 2015; Nair et al., 2015; Baek et al., 2017), and Fig. 3, D-ii and D-iii, are just two examples of microneedles coated in this manner. In yet another method, the approach -distance of microneedles and coating liquid is restricted by placing a physical barrier between them in the form of a wall called the dam, which stops the microneedle base from fully approaching the surface of the coating solution (Chen et al., 2017; Caudill et al., 2018). Using a different approach for creating a thin film, Lee et al. (2017a) used a microneedle array to itself to hold the coating solution such that the liquid filled the valleys in between the microneedles, and then dipped a second microneedle array into these valleys (Fig. 3E-i) (Lee et al., 2017a). An example of a microneedle array coated using this method is shown in Fig. 3E-ii.
Factors affecting dip coating.
Thickness of the coating deposited on microneedle shafts is an important determinant of the mass of drug coated on a microneedle patch; the greater the coating thickness, greater will be the mass coated. Different parameters can be tuned to control the coating thickness, one is the speed at which the microneedle exits the coating liquid A higher microneedle withdrawal speed from the coating liquid will lead to entrainment of a thicker liquid film on the microneedle surface, which will in turn lead to deposition of a thicker coating compared with a coating produced from lower microneedle withdrawal speeds (Gutfinger and Tallmadge, 1965; Scriven, 1988). However, studies investigating the effect of microneedle withdrawal speed are lacking, and most studies have neither controlled nor reported the measure of this parameter. The second parameter that can be tuned to control coating thickness is the solution viscosity (Gutfinger and Tallmadge, 1965). An increase in viscosity of the coating liquid can increase the thickness of the liquid film that is entrained, which can in turn lead to deposition of thicker coatings. For this reason, a viscosity enhancer is often added into the coating solution to increase the viscosity. The third parameter that can be used to control coating thickness is the number of times a microneedle is dipped into the coating solution, the greater the number of dips, the greater will be the coating thickness (Gill and Prausnitz, 2007b; Kim et al., 2010b, 2014; Pearton et al., 2012; Chen et al., 2015; Jain et al., 2016). Additionally, to achieve uniform and nonpatchy coatings, a surfactant is added into the coating solution. The surfactant helps to reduce the surface tension of the coating solution, which in turn permits proper wetting of the microneedle surface (Bierwagen, 1991). This ensures that the entire dipped portion of the microneedle gets coated without formation of uncoated spots. Another factor that can affect the amount coated is the drying time between dips. Pearton et al. (2012) found that increasing the drying time from 5 to 30 seconds significantly increased the mass of DNA coated on microneedles.
Inkjet Printing.
In this method the material to be coated is ejected directly onto the surface of the microneedle in the form of tiny liquid droplets from piezo-driven nozzles (Uddin et al., 2015; Boehm et al., 2016; Pere et al., 2018). Fig. 3F-i shows a piezoelectric nozzle coating a microneedle on an array, and Fig. 3F-ii shows a microneedle array coated with this approach. In contrast to the dip-coating method, for inkjet printing the coating solution viscosity must be relatively low to prevent nozzle clogging and to successfully generate small-sized drops (Uddin et al., 2015). The inkjet printing method can precisely deposit the droplets onto the microneedle surface; however, since the droplets are small, multiple drops maybe required to cover the desired length of the microneedle. Unlike dip coating, inkjet printing is unidirectional, thus at a given time only one side of the microneedle can be coated with a nozzle. A second coating cycle is required if both sides of a microneedle are to be coated. In this coating method, if multiple microneedles are to be coated simultaneously, then multiple nozzles are required. While this is achievable, however, if the microneedle spacing is small, then placement of multiple nozzles with appropriate spacing could become challenging.
Coating Methods Shown to Coat Both the Microneedle Shafts and the Microneedle Array Base Substrate
Immersion Coating.
In this method the entire microneedle array is immersed into a coating solution. Upon withdrawal of the array, the microneedles and the base substrate both get coated (Fig. 4, A-i and A-ii). It should be noted that in this approach, both the back and the front sides of the base of the array also get coated. This was the first technique reported in literature to coat microneedles (Matriano et al., 2002) and has also been used by others (Raja et al., 2013; van der Maaden et al., 2014; Zeng et al., 2017; Chandler et al., 2018). This is the simplest of all coating techniques, but it leads to significant wastage of the active molecule since coatings are produced on the entire microneedle patch.
Drop Coating.
In this method, instead of dipping the microneedles in a coating solution, a drop of the coating liquid is placed on the microneedle array thereby submerging all the microneedles (Fig. 4B-i). This approach prevents coating of the backside of the base substrate but still leads to coating deposition on the front surface of the microneedle array base substrate. As the solvent evaporates, the dissolved solid is expected to conformally coat the microneedles and the base substrate. However, under ambient conditions, the rate of evaporation of the solvent (water in most cases) is not fast enough, and consequently the liquid segregates away from the microneedle tip and accumulates on the substrate between the microneedles. As a result, conformal uniform coatings that cover the microneedle shafts and the base substrate are not achieved (Chen et al., 2009; Vrdoljak et al., 2012) (Fig. 4B-ii). Instead, the coatings are produced primarily on the base substrate in between the microneedle shafts. To address this limitation, the rate of solvent evaporation can be increased by heating the patch (Gittard et al., 2013) or by drying under vacuum (DeMuth et al., 2013). Another approach called gas jet–assisted drying coating has been developed to accelerate the drying process with the expectation that the dissolved solids can produce a conformal coating preferentially on the microneedles (Fig. 4C-i). Chen et al. (2009) first reported this technique and used a gas jet at 6–8 m/s and an incident angle of 20° to the horizontal to direct the coating liquid onto the microneedles and away from the base of the microneedles. The 6–8 m/s jet flow was followed by a faster gas-jet of about 10 m/s to blow away the excess coating solution from the base of the microneedles (Chen et al., 2009). While this coating method can preferentially coat the microneedles, it does not completely prevent the base substrate from being coated (Fig. 4C-ii).
Spray Coating.
Conventional spray coating, which uses fluid pressure as the driving force to create droplets (Fig. 4D-i), has also been used to coat microneedles (McGrath et al., 2011; Vrdoljak et al., 2012). An important consideration in achieving conformal coatings via spray coating is to promote coalescence of the drops deposited on the microneedles and the substrate. It was found that a balance between the viscosity and surface tension of the coating liquid and the droplet size (smaller is better) is essential to achieve conformal coatings (McGrath et al., 2011; Vrdoljak et al., 2012). McGrath et al. found that using this method, conformal coatings were produced on both the microneedle surface and the base substrate (McGrath et al., 2011). In another study, Vrdoljak attempted to spray-coat onto a silicon microneedle patch, and found that instead of a conformal coating, the coatings localized at the base of the microneedles (Fig. 4, D-ii and D-iii) (Vrdoljak et al., 2012). As a variation to the conventional spray-coating method, which uses fluid pressure to form droplets, an electrospraying technique called electrohydrodynamic spray method has also been used to coat microneedles. In this approach electrical potential is used to create droplets. To achieve droplet formation, the microneedle array is grounded while the liquid nozzle is provided a high voltage (Fig. 4E). Khan et al. (2014) coupled this method with a sacrificial insulating mask that contained holes and slid it over the microneedles to protect the base substrate from being coated. Although the insulating mask prevented coating of the base substrate, it did not prevent wastage of drug that is deposited on the mask instead of the base substrate. Readers are referred to a recent review by Haj-Ahmad (Haj-Ahmad et al., 2015), which contains a more detailed explanation of spray coatings.
Layer-by-Layer Coating
Layer-by-layer coating is not necessarily a different method of coating microneedles, it differs rather in the principle by which coatings are produced on microneedles. In layer-by-layer method, instead of relying on solution viscosity, electrostatic interactions are used to create a layered coating on microneedle surfaces. First the microneedle surface is prepared so that it can acquire the desired charge polarity. This is often done by either chemically modifying the microneedle surface (van der Maaden et al., 2015; Duong et al., 2018) or by precoating multiple alternate layers of negatively and positively charged polymers through the standard layer-by-layer coating protocols (Saurer et al., 2010). The charge of the top polymer layer dictates the final charge of the microneedles. Subsequently, alternate layers of drug and a suitably charged material is coated using the layer-by-layer method (Saurer et al., 2010; DeMuth et al., 2012; van der Maaden et al., 2015; Schipper et al., 2017; Zeng et al., 2017). The layer-by-layer protocol typically involves either submerging the entire microneedle array into the solution (DeMuth et al., 2012; Zeng et al., 2017) or placing a coating liquid drop on the array to submerge the microneedles (van der Maaden et al., 2015). The contact time of the liquid drop and microneedle array varies anywhere between a few minutes (Saurer et al., 2010) to hours (van der Maaden et al., 2015). To coat nanoparticles, a spray method has also been used (DeMuth et al., 2012). Some exemplary microneedles coated with layer-by-layer method are shown in Fig. 4, F and G. The layer-by-layer approach cannot be used to coat the microneedles selectively, as the base substrate is also contaminated.
Coating Solution Excipients
To successfully create coatings on microneedles, different coating methods require different values of coating solution viscosities and surface tension. To impart these desirable properties to the coating liquid, different excipients have been added to the coating liquid. The materials that have been used as viscosity enhancers can be classified as synthetic polymers such as polyvinylpyrrolidone (Seok et al., 2017) and polyphosphazene (Andrianov et al., 2009) or natural polymers such as carboxymethylcellulose (Gill and Prausnitz, 2007a), although small molecules such as sucrose (Widera et al., 2006) have also been used. The different materials used as viscosity enhancers are tabulated in Table 1. The surfactants that have been used include polymers such as poloxamer Lutrol F68 (Gill and Prausnitz, 2007a,b) and the more conventional Tween class of molecules (Widera et al., 2006), although some unconventional molecules such as Quil-A have also been used as a dual-use material for its adjuvant effect and surfactant properties (Fernando et al., 2012). The different surfactant materials used in microneedle coatings are summarized in Table 2. Besides adding surfactants to the coating liquid, the microneedle surfaces have also been subjected to different treatments with the goal of changing the microneedle surface energy such that proper microneedle wetting can be achieved with coating liquids without the use of a surfactant. For example, plasma treatment was helpful in reducing contact angle of the coating solution on stainless steel microneedle surfaces (Uddin et al., 2015), and a coating of silicon dioxide on stainless steel microneedles also allowed for an excipient-free coating (Gill and Prausnitz, 2007b).
In addition to viscosity enhancers and surfactants, stabilizers molecule have also been added to help stabilize and protect active molecules such as proteins, inactivated viruses, and virus-like particles from desiccation forces and denaturation. The stabilizer molecules have typically been selected from materials already known for their ability to serve as stabilizers in freeze-drying processes. A study investigated the effect of different excipients and stabilizers and found that instead of carboxymethylcellulose, hydroxyethylcellulose and methylcellulose without surfactant but with trehalose were better able to stabilize hemagglutinin titers in viruses (Kim et al., 2016a). Use of disaccharides such as trehalose, sucrose, and maltose as excipients has helped to stabilize a broad range of materials, such as DNA (Pearton et al., 2012), live measles virus (Edens et al., 2013), erythropoietin (Peters et al., 2012), live adenovirus and modified vaccinia virus Ankara (Vrdoljak et al., 2012), inactivated influenza virus (Quan et al., 2009; Kim et al., 2010a; Kommareddy et al., 2013), inactivated rotavirus (Moon et al., 2013), human growth hormone (Ameri et al., 2014), live bacillus Calmette-Guérin vaccine (Hiraishi et al., 2011), influenza virus-like particles (Kim et al., 2010b), ebola virus glycoprotein (Liu et al., 2018), and inactivated canine influenza virus (Choi et al., 2018). A list of stabilizers with their respective ranges that have been added to coating liquids are presented in Table 3.
Properties to Consider When Selecting Coating Excipients
When selecting a coating excipient for coating microneedles, a few important criteria should be considered. First, the material should be biocompatible, preferably it should already be approved for use as an excipient in injectable formulations by local regulatory agencies, such as the Food and Drug Administration in the United States. Prior approval for use of the coating excipients for injectables can reduce the regulatory hurdle in bringing the coated microneedle product to market. Often the “generally recognized as safe” label is used to motivate the notion that the excipient added to coating formulations is safe. It should be noted though that it applies to food additives and not to injectable excipients. Second, the excipients should be in solid state at room temperature. This is important so that they can get deposited as solids on the microneedle surface when the solvent evaporates. If excipients are in liquid state at room temperature, then the coatings will stay wet even after evaporation of the solvent. Tween 20, which is a viscous liquid at room temperature, has been used as a surfactant by many investigators (Table 2). However, it has been added in low amounts (typically 0.2% w/w), and thus its presence has not adversely impacted the integrity of the coated solid film. Third, the coating excipients should be used in as small an amount as possible. If larger amounts are used, this automatically reduces the fraction of the active component in the coatings, and thus lowers the drug-carrying capacity of the microneedle array.
In Vitro Characterization of Coated Microneedles
Once coatings have been deposited on microneedles, it is important to determine the uniformity of coating, the amount of drug that has been coated, and the delivery efficiency.
Coating Uniformity
A visual analysis of the coated microneedles under a microscope (often a stereomicroscope) can provide details on coating uniformity, and whether coatings are contaminating the base substrate of the microneedle array. This is typically done by adding a colored compound or a fluorescent dye into the coating solution, or by conjugating a fluorescent dye onto the active molecule being coated. This aids in visualization of coatings under a stereomicroscope (Fig. 3, A-iii, D-ii, and D-iii; Fig. 4, A-i and A-ii), or a confocal microscope (Fig. 3B-ii-right; Fig. 4, C-ii and F-i). Alternatively, coated microneedles can also be imaged using a scanning electron microscope (Fig. 3, B-ii-left, Fig. 3, C-ii, E-ii, and F-ii; Fig. 4, D-ii and G).
What Is the “Effective Amount” of Drug That Is Coated on a Microneedle Array?
Since only microneedle shafts penetrate the skin, the drug coated just on the microneedle shafts should be considered as the “effective amount” of drug coated on a microneedle array. For microneedle arrays in which the drug coating is situated and localized on just the microneedles shafts, the drug on the entire microneedle array becomes the “effective drug.” To make this quantification, coated microneedle arrays are often immersed in water to allow for the coating to completely dissolve, and the amount of drug in the solution is then determined. However, when the drug is coated on microneedle shafts and on the base substrate of the microneedle array, it is not feasible to determine the effective drug loading of the microneedle array. This is because once drug from the patch has been dissolved into the extraction liquid, the fraction of drug that emanates from the microneedle shafts cannot be discriminated. In such cases, the amount of drug that is actually delivered into the skin provides a measure of the “effective amount” of drug coated on a microneedle array.
As shown in Table 4, the “effective amounts” of active materials that have been coated on microneedles has ranged from nanograms to hundreds of micrograms to a couple of milligrams. The approximate weight ranges of the different materials that have been coated are: viruses and bacteria, 0.43 ng (Corbett et al., 2010) to 30–45 μg (Kommareddy et al., 2013); virus-like particles, 0.4 μg (Song et al., 2010) to 10 μg (Pearton et al., 2013); oligonucleotides, 4 ng (Kask et al., 2010) to 40 μg (Chong et al., 2013); peptides and proteins, 0.05 μg (Schipper et al., 2017) to 500 μg (Ameri et al., 2014); and small molecules, 132 ng (Chen et al., 2017) to 1.9 and 3.8 mg (Kellerman et al., 2017). Because the dip-coating method provides control and precision in coating, it has been successfully used to coat materials ranging from a few nanograms to milligrams. On the other hand, drop coating and layer-by-layer methods have produced coatings that contain relatively lower amounts of active materials.
Delivery Efficiency
Delivery efficiency is defined as the fraction of the drug coated on a microneedle array that gets delivered into the skin. Although this is an important parameter, during our review we found that not all published studies have determined it. Delivery efficiency is often computed by finding the mass of the drug coated on microneedles, and then subtracting the mass of the drug left on the microneedle array after its insertion into the skin and the mass of the residual drug that maybe left behind on the skin. For microneedles coated with the dip-coating process, the delivery efficiency has been reported in the range of 60%–90% for skin (Hiraishi et al., 2011; Zhang et al., 2012b; Zhao et al., 2016; Baek et al., 2017; Shakya et al., 2017; Nguyen et al., 2018), about 65%–90% for the oral cavity (Ma et al., 2014; Serpe et al., 2016), and about 50%–74% for the eye (Jiang et al., 2007; Kim et al., 2014). Delivery efficiencies from microneedle arrays coated via drop coating (including gas jet–assisted drying coating and layer-by-layer) have been reported to be in the range of 7%–67% for skin (Prow et al., 2010; Fernando et al., 2012; Pearson et al., 2015; Schipper et al., 2017) and about 31% for oral mucosa (McNeilly et al., 2014). For spray coated microneedle arrays, a delivery efficiency of about 40% was achieved (Vrdoljak et al., 2012).
Factors Affecting Delivery Efficiency.
Deposition of coatings only on microneedle shafts.
Since only microneedle shafts penetrate the skin, any drug coated on the microneedle array but away from the microneedle shafts is undeliverable and contributes toward an inefficient delivery process. The contribution of the microneedle array base area toward the inefficiency can be calculated. For example, the typical microneedle patch design used in gas jet–assisted drying coating comprises a 4 × 4 mm patch with 3364 microneedles (Chen et al., 2011). Each microneedle has a base diameter of 25 μm and a length of 110 μm. Although these microneedles are not conical, for an approximate calculation, the microneedles are assumed to be conical in shape, and their total surface area is then about 14.6 mm2 and the surface area of the space between microneedles is about 14.3 mm2. Thus, if drug is uniformly coated on the patch, about 50% of the drug is wasted since it coats the base substrate of the microneedle array. Thus, to achieve higher delivery efficiencies, the base substrate of the microneedle patch should remain uncoated.
Mass of material coated on microneedles.
It has been observed that as the amount of material coated is increased, then after a certain threshold amount, the delivery efficiency decreases (Widera et al., 2006; Kim et al., 2014; Jain et al., 2016; Abdalla et al., 2019). In a study done by Widera et al. (2006) it was seen that the delivery efficiency in hairless guinea pig skin using microneedles coated with radiolabeled ovalbumin was a function of the amount coated. A delivery efficiency of 27% was observed when a high dose (25 μg) was coated; however, the delivery efficiency increased to 50% when a lower dose (5 μg) was coated on the same patch. Likewise, Jain et al. (2016) found that although the amount of 5-aminolevulinic acid on a microneedle patch increased from 206–350 to 458–680 μg/patch, the delivery efficiency was found to have fallen from about 90% to 60% across the dose range. Abdalla et al. (2019) also found a similar threshold effect. When they coated tramadol onto microneedles, they saw that at 20% or 30% (w/v) of tramadol in the coating solution, the resulting delivery efficiencies were similar and measured at 73% and 72%, translating into delivery of 139 and 185 μg tramadol, respectively. However, when the mass of tramadol in the coating solution was increased to 50% (w/v), then, despite the presence of a higher mass of tramadol in the coatings, the delivery efficiency fell to 26% and a much lower mass of tramadol got delivered (78 μg). A similar phenomenon was observed upon insertion of microneedles into the cornea of anesthetized rabbits (Kim et al., 2014). It was seen that although the amount of bevacizumab coated on microneedles increased from 1.1 to 7.6 μg, the delivery efficiency was reduced from around 52% to 44%.
Residence time of coated microneedles in tissues.
It has also been observed that if the coated amount is fixed but the wear time of the patch is increased, the delivery efficiency increases (Cormier et al., 2004; Baek et al., 2017; Lee et al., 2017a). In a study done by Peters et al. (2012) a delivery efficiency of 75% was achieved when a patch coated with 100 μg of erythropoietin was placed in skin for 15 minutes and was further enhanced to 90% when the patch was placed for 2 hours. In another study, Zhang et al. (2012a) coated lidocaine on microneedles and observed a similar effect. The delivery efficiency for a lower dose (approx. 90 μg/array) increased from 53% to 71% as the wear time was increased from 1 to 4 minutes. At a higher dose (approx. 225 μg/array), a much lower delivery efficiency was seen after a 1-minute wear time (approx. 16%), which increased to 28% with a wear time of 4 minutes; however, it was still significantly lower compared with the low dose group. One possible explanation for this phenomenon is that if a larger amount of material is delivered into the skin using coatings, then the material does not immediately dissolve in the limited local interstitial fluid available in the tissue. As a result, if the microneedles are removed while there is still undissolved material at the deposition site, then the microneedle can pull back a fraction of this material, leading to lower delivery efficiencies. However, if the microneedle is allowed to stay inserted in the skin, then over time, the material dissolves in the local tissue fluid and dissipates away from the delivery site.
Other factors.
Delivery efficiency can also be affected by hydrophobicity of the active molecule. In one study, keeping all parameters same, use of low, medium, and high water-soluble peptides led to significantly different delivery efficiencies of about 46%, 59%, and 90%, respectively (Zhao et al., 2017). Delivery efficiency can also vary on the basis of the site of insertion. For example, it was observed by Ma et al. (2014) that using microneedles coated with ovalbumin, a delivery efficiency of 64% was achieved in rabbit lip, whereas a significantly higher delivery efficiency of 91% was observed when the patch was inserted in the tongue. The flat area available to insert a microneedle patch is comparatively smaller in the lip compared with the tongue, which could have reduced the delivery efficiency.
Material of Microneedle Fabrication
A variety of materials have been used to make the coated microneedles. The most dominant material reported in literature is stainless steel, followed by silicon, titanium, and other materials, including polymers. Since microneedles are often dipped in aqueous solutions for coating, they should not be constructed from water-soluble materials. A few studies have, however, reported microneedles made from hyaluronic acid (Liu et al., 2016; Choi et al., 2018). In these studies, to coat microneedles made from hyaluronic acid, the microneedles were dipped rapidly into the aqueous coating solutions to minimize microneedle dissolution (Liu et al., 2016; Choi et al., 2018). In one study however, the microneedles were first frozen to reduce their solubility in water (Choi et al., 2018).
Biomedical Applications, Animal Models, and Human Studies Pursued with Coated Microneedles
Not only is the skin an attractive anatomical site for vaccine delivery, but it also provides an attractive pathway for the delivery of other active compounds. Drugs delivered via coated microneedles bypass the hepatic first-pass effect and can also eliminate adverse effects such as gastric irritation, which can arise if drugs are delivered orally. A broad range of active materials have been coated on microneedles. These include small molecules, peptides, proteins, short nucleic acids, DNA, viruses, virus-like particles, polymeric particles, and insoluble inorganic particles (Cormier et al., 2004; Gill and Prausnitz, 2007a; Kim et al., 2010b; Zhang et al., 2012b). More specifically, coated microneedles have been widely investigated to study influenza vaccination (Kim et al., 2011; McNeilly et al., 2014; Wang et al., 2014; Fernando et al., 2016; Choi et al., 2018). Additionally, coated microneedles have also been used to deliver antigens for other infectious diseases such as diphtheria (Schipper et al., 2017; Du et al., 2018), ebola (Liu et al., 2018), hepatitis (Andrianov et al., 2009; Gill et al., 2010), herpes simplex virus (Chen et al., 2010; Kask et al., 2010), human immunodeficiency virus (Ma et al., 2014; Caucheteux et al., 2016), human papillomavirus (Kines et al., 2015), Leishmania (Moreno et al., 2017), measles (Edens et al., 2013), pneumonia (Pearson et al., 2015), polio (Muller et al., 2016), severe fever with thrombocytopenia syndrome (Jung et al., 2017), smallpox (Hooper et al., 2007), tuberculosis (Hiraishi et al., 2011), West Nile virus (Prow et al., 2010), and chikungunya (Prow et al., 2010). Microneedles have also been used for treatment of allergy (Shakya and Gill, 2015), local delivery of anesthetics (Zhang et al., 2012a), for treatment of anemia (Peters et al., 2012), treatment of type 2 diabetes (Liu et al., 2016), treatment of intimal hyperplasia (Lee et al., 2017b), corneal neovascularization (Kim et al., 2014), treatment of enuresis (Cormier et al., 2004), and treatment of cancer (Ma et al., 2015; Jain et al., 2016). Coated microneedles have also been used in immunotherapy of type 1 diabetes (Zhao et al., 2017) and for treatment of osteoporosis (Katsumi et al., 2017) and restenosis (Lee et al., 2014).
In addition to mice and rats, larger animals whose skin better resembles that of humans have also been used for in vivo investigation of coated microneedles. These larger animals include guinea pigs/hairless guinea pigs (Kommareddy et al., 2013), pigs (Andrianov et al., 2009; Zhang et al., 2012b), and macaques (DeMuth et al., 2013). Rabbits have also been used to deliver drugs and vaccines into the eye (Jiang et al., 2007) and the mouth (Ma and Gill, 2014).
Coated microneedles have been tested in humans. For example, coated microneedles have been used to study treatment of warts (Ryu et al., 2018), osteoporosis (Daddona et al., 2011), migraine (Spierings et al., 2018), and influenza vaccination (Fernando et al., 2018). The tolerability and acceptability of excipient-coated microneedles in humans has also been evaluated. Griffin et al. (2017) noted redness of the skin (erythema) in humans, which lasted day 3 to day 7 post–microneedle application. Pin-prick bleeding was observed in some subjects but it lasted only until 10 minutes post-application. It was found that most of the subjects preferred microneedle patch application over needle and syringe (Griffin et al., 2017). A safety study (Ono et al., 2017) and pain evaluation from uncoated microneedles (Gill et al., 2008) has also been done in humans.
Factors to Consider When Deciding Suitability of Coated Microneedles for New-Drug Delivery Applications
When deciding whether coated microneedles are a suitable drug delivery system for a new application, the drug physical properties at room temperature, drug potency, and drug water solubility are some of the important factors to consider. For obvious reasons, drug that is in a liquid state cannot be delivered using solid microneedles since the liquid will wipe off on the skin surface during insertion. The physical state of the drug at room temperature must be a solid so that when the solvent in the coating solution evaporates, the drug that remains behind can produce a solid film on the microneedle shaft. One of the most important questions to ask in deciding suitability of coated microneedles for drug delivery is the amount of drug that is to be delivered. This is because the delivery capacity of coated microneedles is limited. In general, a microneedle patch of about 1 cm2 can carry as much as a few hundred micrograms of drug. Thus, potent drugs are a good candidate for delivery via coated microneedles. However, if multiple microneedle patches are used or the microneedle patch size is increased, then much higher amounts of drug can be delivered. For example, in a human study, 3.8 mg of drug (zolmitriptan) was delivered using two 5-cm2 patches each coated with 1.9 mg of drug, and also with a single 10-cm2 patch that was coated with 3.8 mg, and the drug pharmacokinetics were found to be similar (Kellerman et al., 2017). Another factor to consider is the water solubility of the drug. Drug water solubility is important because the coating solution often has water as the solvent, and thus it is important that the drug should be water soluble to allow its dissolution in the coating liquid. However, it should be noted that materials such as viruses, bacteria, nanoparticles, and microparticles, although not water-soluble, can nevertheless form a stable dispersion in water and thus be coated on microneedles. These dispersions do not have to be indefinitely stable but should stay stable at least during the coating process. Although it is preferable that drugs should be water soluble, even hydrophobic materials can be coated on microneedles by creating special coating formulations, such as by using cosolvents for dissolving the material in the coating solution (Zhao et al., 2017). The drug stability is an important factor, which is especially relevant to proteins, and must be considered when making a decision to pursue drug delivery using coated microneedles.
Conclusion and Perspective on Microneedle Coatings
Of the different coating methods, some are better than others at achieving the two important properties desirable in a coating process, i.e., achievement of uniformity in coating and the ability to selectively coat just the microneedle shafts without coating the microneedle array base substrate. The dip-coating methods and the inkjet-coating method are precision coating methods that coat just the microneedle shafts without causing wasteful depositions on the substrate. The gas jet–drying coating version of drop coating has improved over the years and offers better selectivity in coating the microneedle shafts; however, its delivery efficiencies are still lower than the microneedles coated with the dip-coating approaches. The spray-coating method indiscriminately sprays the drug on the microneedle patch, and although a sacrificial mask can be used to prevent the coating of the substrate of the microneedle, the drug that gets coated on the sacrificial mask is still wasted. In the layer-by-layer method, selective coating of just the microneedles is not achieved; instead, the coatings are also deposited on the base substrate. Thus, the layer-by-layer coating approach also inherently suffers from poor drug delivery efficiencies. However, if the dip-coating approach is used for layer-by-layer coatings, higher drug delivery efficiencies could be achieved. In terms of time taken to coat a microneedle array, the layer-by-layer method on average is expected to take longer than the other coating methods.
Delivery efficiency of a coated microneedle is an important and biologically relevant design parameter because it directly affects the dose of the active material that is administered in vivo. The delivery efficiencies of microneedle arrays coated via the dip-coating method average around 70%–90%, but this number is reduced to around 30%–50% for immersion coating, drop-coating, spray-coating, and layer-by-layer coating methods. Delivery efficiency is a critical parameter, however, during our review of the literature, we found that many studies failed to report delivery efficiency of their coated-microneedle array. To facilitate cross-study comparisons and to promote rigor, the quantification and reporting of delivery efficiencies of coated-microneedle patches should be considered essential. It should be noted that quantification of proteins and oligonucleotides on microneedles after their insertion into tissues, via nonspecific assays that rely on quantifying total proteins and oligonucleotides can lead to erroneous results and should be avoided. This is because when the coated microneedles are removed from tissues, they can transport out native proteins and DNA from the tissues, and these molecules can erroneously increase the test readings. Fluorescent or radiolabeled tags or tracer molecules are better approaches to facilitate evaluation of delivery efficiencies (Widera et al., 2006; Crichton et al., 2010; Hiraishi et al., 2011; Chen et al., 2012; Luo et al., 2013). Inherent spectroscopic properties of the active material have also been used to specifically quantify the active materials (Peters et al., 2012; Ameri et al., 2014; Jain et al., 2016).
Coated microneedles offer a versatile approach for the delivery of therapeutics. Importantly, different researchers have shown that by using the same base microneedle structure and with almost no change in coating excipients, a few nanograms to hundreds of micrograms of a broad assortment of active materials can be coated including small molecules, proteins, DNA, viruses, and micro- and nanoparticles. For example, in our own coating studies at Texas Tech University (Lubbock, TX), we have continued to use carboxymethylcellulose (1% w/v) as the viscosity enhancer and Lutrol F68 (0.5% w/v) as the surfactant in all our experiments. Likewise, the Prausnitz laboratory at Georgia Institute of Technology (Atlanta, GA) has also been using these same excipients for most of their studies. They have also added trehalose at varying amounts to confer stability on the more labile active materials that are at risk of denaturation upon drying. The work coming out of 3M and Zosana Pharma, likewise, have used sucrose in varying amounts as the viscosity enhancer, with Zosana Pharma also using polysorbate 20 at 0.2% (w/w) as the surfactant. The Kendall group at the University of Queensland (St. Lucia, Australia) has chiefly used methylcellulose as the viscosity enhancer for coating a broad range of materials. The significance of this observation is that when faced with a situation of coating a new active material, the same coating formulation without modification can often be used as a good starting point, and the targeted mass in the coating can be achieved by simply changing the weight fraction of the active material in the coating solution. This suggests that not only the same base microneedle array but also the same coating formulation could be used to develop new products based on coated microneedles, which can help to accelerate product development.
Recently, the concept of individually coated microneedles (Caudill et al., 2018; Li et al., 2018) has emerged as a means of coating different drug molecules on the same microneedle patch. Using this concept, it should be possible to provide combination drug therapies through a single microneedle patch. In addition to their use in the delivery of active molecules, coated microneedles are now being developed for use in sampling the skin and the interstitial fluid for detection and sensing purposes (Mandal et al., 2018; Samant and Prausnitz, 2018).
In addition to water-based coating solutions, a molten coating solution without a solvent has also been reported. To coat water-insoluble drugs, Ma and Gill (2014) used a mixture of polyethylene glycol (PEG)+lidocaine (water-insoluble base form of lidocaine), heated it to 120°C to melt the PEG and lidocaine, and dip-coated the microneedles in it. PEG was added as an excipient to serve as a water-soluble matrix so that upon insertion into skin PEG would dissolve and facilitate detachment of the PEG+lidocaine coating from the microneedle surface. Such approaches could be used to develop formulations for coating water-insoluble small-molecule drugs onto microneedles, which could increase the scope of materials that can be delivered using coated microneedles.
From a commercialization viewpoint, four companies, namely Zosana Pharma (https://www.zosanopharma.com/, accessed: April 1, 2019), 3M (https://www.3m.com/3M/en_US/drug-delivery-systems-us/technologies/microneedle/solid/, accessed: April 1, 2019), Vaxxas (http://www.vaxxas.com/, accessed: April 1, 2019), and Moonlight Therapeutics (https://www.moonlighttx.com/, accessed: April 1, 2019) are operating in the coated-microneedle space. Zosana Pharma and Moonlight Therapeutics use metal microneedles as their base platform, 3M uses polymer microneedles, and Vaxxas use silicon microneedles. These four companies are seeking to develop products with coated microneedle technology as a basis. Zosana Pharma, 3M, and Moonlight Therapeutics use the dip-coating approach to coat microneedles, whereas Vaxxas uses the drop gas jet–assisted drying coating method. Zosana Pharma has a coated microneedle patch for treatment of migraine, and they have completed a phase 3 trial in humans. Vaxxas is developing coated microneedles for vaccine delivery and for use in oncology. 3M on the other hand has developed capability to coat microneedles and works directly with partners for custom applications. Moonlight Therapeutics is developing coated microneedles for treatment of allergies and is focusing on peanut allergy as their first product.
In conclusion, a wide range of microneedle coating methods have been developed. For most applications, controlled coating of microneedle shafts without contaminating the base substrate is desirable. It can, however, be argued that in some instances, e.g., when the drug is inexpensive, contamination of the base substrate may be tolerated. So far, the dip-coating and inkjet-coating methods have exhibited most promise in facilitating the selective coating of microneedle shafts without contaminating the base substrate. The most widely studied use of coated microneedles has been for the delivery of infectious disease vaccines; however, other applications are being actively investigated. Although additional fundamental research on the coating processes can benefit the field of coated microneedles, there is now a strong effort, after 17 years from when the first article on coated microneedles was published (Matriano et al., 2002), to bring the first coated microneedle–based product to market.
Authorship Contributions
Participated in research design: Ingrole, Gill.
Performed data analysis: Ingrole, Gill.
Wrote or contributed to the writing of the manuscript: Ingrole, Gill.
Footnotes
- Received April 5, 2019.
- Accepted May 31, 2019.
This publication was supported by the National Institute of Allergy and Infectious Diseases National Institutes of Health [Award Numbers R01AI135197 and R01AI121322 to H.S.G.]. The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIH. Financial Disclosures: HSG is a coinventor on a patent related to coated microneedles and has stock ownership in a startup company called Moonlight Therapeutics that is developing microneedles for food allergy immunotherapy. This conflict of interest has been disclosed and is being managed by Texas Tech University. No financial support was provided by Moonlight Therapeutics for this work.