TABLE 1

HPLC and LC/MS/MS methods


Parameter

HPLC Method 1

HPLC Method 2

LC/MS/MS
Apparatus Agilent 1100 HPLC system with binary pump, auto sampler, column thermostat, and diode array detector. Waters Alliance HT 2790 coupled to a Micromass Quattro Ultima II triple quadrupole mass spectrometer.
Solvent A 0.02% HFBA in water (v/v) 0.1% TFA in water (v/v) 0.02% HFBA in 5:95 MeOH/water (v/v/v)
Solvent B 0.02% HFBA in 10:90 water/MeOH (v/v/v) 0.1% TFA in 10:90 water/MeOH (v/v/v). 0.02% HFBA in 10:90 water/MeOH (v/v/v)
Column Luna C18(2), 150×2 mm, 3 μm (30°C) (Phenomenex, Torrance, CA) Vydac 218MS52 C18, 250×2.1 mm, 5 μm (30°C) (Vydac, Hesperia, CA) XterraMS C18, 50×3 mm, 3.5 μm (50°C) (Waters, Milford, MA)
Flow rate 0.200 ml/min 0.200 ml/min 0.200 ml/min
Gradient Initial: 5% B Initial: 0% B Initial: 0% B
Linear gradient: 0-5 min, 5 to 30% B Linear gradient: 2-14 min, 0 to 25% B Linear gradient: 0-4 min, 0 to 100% B
Isocratic: 5-15 min, 30% B Wash: 15-16 min, 100% B Wash: 4-4.9 min, 100% B
Wash: 16-17 min, 95% B Equilibration: 17-30 min, 0% B Equilibration: 5-10 min, 0% B
Equilibration: 18-35 min, 5% B
Detection UV at 214.5 nm UV at 214.5 nm Parent > daughter m/z, (collision energy):
ZP123: 618.10 > 413.15 (15 eV)
AAP10: 576.30 > 391.30 (20 eV)
Capillary and Cone Voltage: 3,000 and 45 V
Cone Gas (N2): 100 l/hr (100°C)
Desolvation Gas (N2): 600 l/h (350°C)
I.S.: 624.1 > 419.15 (15 eV)



Collision gas: Ar (1 × 10−3 mbar)