![]() |
|
|
Vol. 290, Issue 2, 694-701, August 1999
Department of Anaesthesia and Intensive Care, Royal Adelaide Hospital, University of Adelaide, Adelaide, Australia (R.N.U., Y.F.H., D.J.D.); and Department of Anaesthesia and Intensive Care, Flinders Medical Center, Flinders University of South Australia, Bedford Park, Australia (L.E.M.)
| |
Abstract |
|---|
|
|
|---|
The myocardial kinetics of meperidine and the relationship between these kinetics and the effect of meperidine on myocardial contractility (maximum positive rate of change of left ventricular pressure) were examined by analysis of previously published data collected in sheep after the i.v. injection of 100 mg of meperidine over 1 s. There was significant hysteresis between reductions in myocardial contractility and the arterial concentrations of meperidine, but not the coronary sinus blood (effluent from the heart) or calculated myocardial concentrations. The peak reduction in contractility occurred after the peak arterial concentration, at the time of the peak myocardial concentration, but before the peak coronary sinus concentration, suggesting that the site of drug action in the heart was not in equilibrium with either arterial blood or effluent blood from the heart. The most appropriate form of a dynamic model (a linear model with a threshold) was determined, without the need to assume a kinetic model, by directly fitting the observed reductions in myocardial contractility to the calculated myocardial concentrations. To determine the optimal kinetic and combined kinetic-dynamic models, a variety of one-, two-, and three-compartment models of the myocardium were fitted to the coronary sinus concentrations by using hybrid modeling. These included "tank in series" models that accounted well for drug dispersion and "peripheral compartment" models that accounted well for deep distribution. The most appropriate model was a "compilation" model, which incorporated features of both these extremes and was a better fit to the observed data than either a traditional single flow-limited compartment or a traditional membrane-limited model.
| |
Introduction |
|---|
|
|
|---|
Previous
reports have documented experimental data regarding the myocardial
kinetics of meperidine as determined by measurements of the product of
its arteriovenous concentration difference across the heart and
myocardial blood flow after bolus administration in sheep (Huang et
al., 1994a
). In the same experiments, it also was shown that meperidine
[at doses that are sufficiently low to avoid overt central nervous
system (CNS) stimulation] caused transient reductions in myocardial
contractility (Huang et al., 1994b
). This was taken to be a direct
effect on the heart, as reductions in contractile performance also have
been documented in isolated heart tissues without CNS innervation
(Rendig et al., 1980
) and appears to be related to the action of
meperidine on calcium currents rather than opioid receptors (Wu et al.,
1997
).
As part of developing a physiological model of meperidine disposition,
these previously published data were examined to determine an optimal
model of the myocardial kinetics of meperidine and its dynamic effect
on myocardial contractility. The general strategy used first was to
examine the kinetic-dynamic relationship by using model-independent
methods based on the analysis of hysteresis loops. Previously, this
approach has been used for lidocaine in the heart, with mass balance
principles (Upton et al., 1988
) used to calculate the lidocaine
concentration in the myocardium. It was found that the reduction in
myocardial contractility caused by lidocaine was better related to this
myocardial concentration than to its concentration in either arterial
or coronary sinus (effluent from the heart) blood (Huang et al., 1993
).
Second, an optimal dynamic model (e.g., linear, maximum effect
(Emax), or sigmoid
Emax) was determined by directly
fitting the observed reductions in myocardial contractility to the
calculated myocardial concentrations, a method that required no
assumptions about the underlying myocardial kinetics of meperidine.
Third, the myocardial kinetics were studied in detail by determining
the best fit to the data of a variety of one-, two-, and
three-compartment models of myocardial kinetics (including a
traditional single flow-limited compartment and a traditional
membrane-limited model). These were also combined with the optimal
dynamic model to produce a variety of kinetic-dynamic models. The
optimal models were chosen based on hybrid modeling of the time courses
of the coronary sinus concentrations of meperidine and changes in
contractility together with deductions from the model-independent analysis.
There are some outstanding issues with respect to the modeling of
myocardial kinetics and dynamics. It is not clear from the present
literature whether multicompartment models of the myocardium can be
resolved from data regarding the arteriovenous concentration difference
across the myocardium, as collected previously for meperidine.
Furthermore, it is not known whether a more realistic description of
myocardial kinetics can be achieved by adding the extra compartments in
series to represent dispersion (Roberts et al., 1988
) or by adding the
extra compartments as "peripheral" or "deep" compartments as
used in traditional membrane-limited models (Gerlowski and Jain, 1983).
Finally, when considering the relationship of the model to simultaneous
effect measurements in the heart, it is clear that the optimal model is
one in which at least one of the compartments is in pseudoequilibrium
with the site of drug action responsible for the effect. Clearly,
increasing the number of compartments used to describe the myocardium
provides greater flexibility when constructing such a combined
kinetic-dynamic model. The aims of the present study were to explore
these issues by examining our previously published data on the
myocardial kinetics and dynamics of meperidine. The specific aims were
to:
1. determine whether the time course of the effect of meperidine on myocardial contractility was better related to the time courses of its concentration in arterial blood or coronary sinus blood or its calculated mean concentration in the myocardium;
2. determine the optimal dynamic model describing the relationship between the calculated myocardial concentration of meperidine and changes in myocardial contractility;
3. determine whether data regarding the arteriovenous concentration difference of meperidine across the myocardium have sufficient information to distinguish between a variety of compartmental models of myocardial kinetics and, if so, determine which model is optimal for describing the data; and
4. determine whether such models also can describe the effects of meperidine on myocardial contractility.
The general principles elucidated here may be applicable to other models of drug kinetics and dynamics in individual organs.
| |
Materials and Methods |
|---|
|
|
|---|
Data Source
The data on the effect of meperidine on myocardial contractility
and other cardiovascular parameters (Huang et al., 1994b
) and on its
blood and myocardial pharmacokinetics (Huang et al., 1994a
)
after bolus i.v. injection in adult female sheep have been reported
previously. In this paper, the relationship between myocardial contractility and myocardial kinetics is examined for the first time. A
summary of the study design and the experimental methods relevant to
this paper is presented here for completeness.
Adult Merino sheep were prepared under anesthesia with chronic intravascular catheters in the aorta (for sampling afferent blood to the heart), the coronary sinus (for sampling pure efferent blood from the heart once the hemiazygous vein is ligated), and the right atrium (for drug administration). An ultrasonic Doppler flow probe was placed on the left main-stem coronary artery to measure myocardial blood flow and was calibrated ex vivo. A pressure transducer-tipped catheter was acutely placed in the left ventricle (LV) on experimental days via a chronic introducer catheter, and the pressure trace was integrated digitally to give the maximum positive rate of change of LV pressure (LV dP/dtmax), an index of myocardial contractility.
On an experimental day, conscious sheep were administered 100-, 200-, or 300-mg i.v. doses of meperidine over 1 s. Myocardial blood flow
and LV dP/dtmax were recorded continuously for
the next 10 min and at 15 min, while paired arterial and coronary sinus
blood samples were taken at intervals of as short as 5 s for 15 min after the dose. Meperidine blood concentrations were determined by
using a gas chromatographic method. The myocardial concentrations of
meperidine were calculated by using mass balance principles (Huang et
al., 1994a
; Upton, 1994
). By this method, the mass of meperidine in the
myocardium at a given time is the integral of the difference in the
flux of meperidine entering the myocardium [i.e., myocardial blood
flow (Qh) times the arterial meperidine concentration,
Cart] and the flux leaving the
myocardium (i.e., Qh times the coronary sinus meperidine
concentration, Ccs). The myocardial
concentration (Cmyo) is this mass
divided by the mass of the myocardium perfused by the left main
coronary artery (Mmyo), estimated
previously (Huang et al., 1994a
):
|
Data Analysis
Hysteresis Analysis. For the model-independent analysis, the time course of changes in LV dP/dtmax (expressed as percent reduction from baseline) were plotted against the time courses of the arterial, coronary sinus, and calculated myocardial concentrations of meperidine. To quantify the hysteresis, each plot was divided into two sections at the point of the maximum meperidine concentration, with one section representing concentration-effect relationships when meperidine concentrations were increasing and the other representing the relationships when the meperidine concentrations were decreasing. The area under the curve (AUC) of each section of the concentration-effect loop were calculated by using the trapezoid rule. Concentration-effect hysteresis was considered to be present when the differences between the AUC of these two sections of the hysteresis loop were statistically different. The times of the peak values of the concentrations and percent reduction in contractility also were compared.
Optimal Dynamic Model.
To determine the optimal dynamic
model, the relationship between the calculated concentration of
meperidine in the myocardium and the contractility data was analyzed
directly by using Scientist for Windows (Micromath Scientific Software,
Salt Lake City, UT). The fit of linear, linear with a threshold,
Emax, and sigmoid Emax dynamic relationships to these
data were examined by using a least-squares method. The equations for
all but the linear with a threshold models are in common usage and have
been reported previously by our laboratory (Huang et al., 1998
). The
equations of the threshold model (in a form common to many programming
languages) were as follows, where Cmyo
is the calculated myocardial concentration, "dpdt" is the percent
reduction in LV dP/dtmax from baseline, and
"dpdtemp" is a temporary variable:
|
|
|
Optimal Kinetic and Combined Kinetic-Dynamic Models.
Models
of myocardial kinetics with a maximum of three compartments were
considered. To illustrate the possible configurations, consider methods
for improving the fit of a single compartment model by adding a second
compartment
two compartments either may be arranged "in series" to
improve the description of the dispersion process (Roberts et al.,
1988
) or as a more traditional "membrane-limited" model to account
for deep distribution (Gerlowski and Jain, 1983). When three
compartments are considered, it is clear that a three-"tank in
series" model, a "double" membrane-limited model, and other arrangements that combine the features of both of these extremes are
possible. The resultant nine kinetic models are shown in Figs. 1, 2, and
3; the equations describing the models
are shown in Appendix 1.
|
|
|
5
to 104 liters, as exceedingly small volumes had a
minimal contribution to the model but significantly delayed the
curve-fitting process and exceedingly large volumes (always in a
"terminal" compartment) can be underdetermined but are indicative
of an apparent first-order loss from the compartment; this, therefore,
was not recorded as a nonidentifiable parameter.
Combined kinetic-dynamic models were examined by using an extension of
the same process. The concentration in each compartment in the model
was, in turn, linked to the optimal form of the dynamic model
determined previously, and curve-fitting was used to estimate the
kinetic and dynamic parameters of the model from both the coronary
sinus and myocardial contractility data.
Statistical Analysis
Data in the text are presented as mean (lower to upper 95%
confidence intervals), assuming a t distribution (Gardner
and Altman, 1989
). Two means were considered significantly different at
the 95% level if each of the means lay outside of the confidence
intervals of the other or if the 95% confidence intervals of the
difference did not include zero.
| |
Results |
|---|
|
|
|---|
Hysteresis.
The hysteresis loops of effect versus the various
concentrations are shown in Fig. 4. There
was a pronounced counter-clockwise hysteresis for the arterial
concentrations, a smaller clockwise hysteresis for the coronary sinus
concentrations, and an again smaller clockwise hysteresis for the
calculated mean myocardial concentrations. The mean size of the
hysteresis loops (and 95% confidence intervals) as given by the
difference between AUC under the sections of the plots when the
concentrations were increasing and decreasing was 449 mg s
l
1 (242-655) for the arterial
concentrations. This statistically significant hysteresis is clearly
evident in Fig. 4A. Figure 4, B and C, shows the AUC differences of 37 mg s l
1 (
39 to 113) for the coronary sinus
concentrations and 69 mg s l
1 (
12 to 150) for
the calculated myocardial concentrations; these later two sites were
not statistically different from zero, suggesting no significant
hysteresis.
|
Optimal Dynamic Model.
The MSC of fits of the dynamic models
to the data are shown in Table 1. Both
the Emax and sigmoid
Emax models were underdetermined for
the value of Emax leading to
nonidentifiability for this parameter. Although it could be argued that
using higher meperidine doses would allow the measurement of
Emax, this is physically impossible for myocardial contractility in vivo. The maximum reduction in contractility is the balance between direct depressant effects and
indirect CNS effects, as well as the risk of fatality. The linear model
was a reasonable fit, but returned values for the slope of 3.12, S.D.
0.28, and an intercept of
7.53, S.D. 1.92. This negative
intercept suggests a "threshold" effect for measurable contractility changes, which is broadly consistent with Fig. 4C. The
final model chosen, therefore, was the linear model with a threshold
such that at concentrations below the threshold the reductions in
contractility were zero. This form of the dynamic relationship (but not
the parameter values) therefore was used for the subsequent combined
kinetic-dynamic models, which enabled a considerable reduction in the
number of combined kinetic-dynamic that needed to be examined in the
following section.
|
Optimal Kinetic and Combined Kinetic-Dynamic Models. The MSC of the various models are shown in Table 2. The parameters of the kinetic models and their S.D. returned by the fitting program are given below. Model 1, the single flow-limited compartment, was the worst description of both the kinetic and combined kinetic and dynamic data sets. The general problem was a poor account of the time of the peak and "washout" coronary sinus concentrations (Fig. 5). The volume of the heart (Vh) was 0.73 S.D. 0.02 liter, which equated to an equilibrium half-life between arterial and coronary sinus blood of 4.13 min.
|
|
16 S.D. 6.6 × 10
15. However, when used for the
kinetic-dynamic data, the fit of this model was improved significantly,
but with an underdetermined value for
Vh.
The "triangular" model (model 8) was a reasonable fit of the
kinetic data and had the highest MSC when applied to the
kinetic-dynamic data. For the kinetic data, the value of
V1 was 0.061 S.D. 0.116 liter,
V2 was 0.33 S.D. 0.06 liter,
Vh was 0.06 S.D. 0.08 liter, and
Q2 was 0.12 S.D. 0.03 l/min. Its shortcomings were
underdetermined values for V1 or
Vh, which were correlated with a
correlation coefficient of >0.99.
The double membrane-limited model in the B configuration (model 9) was
not a good description of the kinetic or kinetic-dynamic data and was
underdetermined for many parameters. For the kinetic data,
V1 was constrained to the minimum
value of 10
4,
V2 was constrained to the maximum
value of 105, Vh
was 0.56 S.D. 0.05 liter, PS1 was 0.63 S.D. 0.37 l/min, and PS2 was
6.1 × 10
16 S.D. 1.2 × 10
14.
Summary of Model Performance. Note that across the range of models a good fit to the kinetic data did not necessarily mean a good fit to the kinetic-dynamic data, and vice versa. This is indicative of the fact that a good fit to the dynamic data generally arose when a compartment of the model linked to the dynamic model was "upstream" of the coronary sinus compartment. This is consistent with the model-independent analysis, which also showed that the effect preceded the coronary sinus concentrations.
By examining Table 2 and neglecting those models with nonidentifiable parameters, the rank order of the best kinetic models (from high to low MSC) was models 5, 6, 2, and 1. For the kinetic-dynamic data, the ranked models were models 2, 5, and 1. Of the three models suitable for both data sets, the mean MSC for both data sets was 3.63 for model 5, 3.54 for model 2, and 2.84 for model 1. Therefore, both models 5 and 2 were good at describing these data (although with a slight advantage to model 5) and had in common two compartments in series providing an element of dispersion to the model output. Interestingly, although model 1 (the single flow-limited compartment) fell well short of these two models in describing the data, it performed better than many more complicated models, in agreement with its widespread use in many physiological pharmacokinetic models.| |
Discussion |
|---|
|
|
|---|
An initial observation from this analysis is the general utility of the mass-balance method for describing this type of regional kinetic-dynamic data. In this case, it was used to determine whether the mean myocardial concentration of a drug was related to its proposed effect on the myocardium and was able to do so even if there was incomplete equilibrium between the effect and the venous concentrations emerging from the organ. This itself is good evidence that a single-compartment description of the organ may not be appropriate, as this model assumes instantaneous equilibration between tissue and effluent venous drug concentrations.
A general picture also is emerging for a variety of drugs that changes
in myocardial contractility caused by the drug are better related to
its myocardial rather than arterial concentrations; this has been
demonstrated for thiopental (Upton et al., 1996
), lidocaine (Huang et
al., 1993
), verapamil (Huang et al., 1998
), and, now, meperidine in the
present study. The mass-balance method also was useful for determining
the optimal dynamic model independent of any assumptions about the
underlying myocardial (or effect-compartment) kinetics. It is not clear
whether the observed threshold effect of meperidine on myocardial
contractility was a real effect or was due to lack of sensitivity in
the method. Although the error bars in Fig. 4 are relatively small,
such a threshold effect has not been observed in vitro (Wu et al.,
1997
).
The opportunity was taken in this study to address a pertinent issue
regarding the choice of kinetic models for individual organs or regions
of the body such as the heart. In many physiological models of
lipophilic drugs such as meperidine (Gabrielsson et al., 1986
; Davis
and Mapleson, 1993
), the heart and many other organs or regions are
represented as single flow-limited compartments. In agreement with the
present data, it was found that the myocardial kinetics of fentanyl and
alfentanil after a 1-min infusion to rats were better described by a
single flow-limited compartment (equivalent to the present model 1)
than either two- or three-compartment models equivalent to models 3 and
7 of the present paper, respectively (Bjorkman et al., 1994
). However,
although single flow-limited compartment models are in broad agreement
with the available data, there is some evidence that this
single-compartment representation of an organ or region is too simple,
particularly for data arising from bolus or impulse administration
studies. For example, when drug kinetics in organs or regions are
examined individually after impulse administration, it generally is
noted that the concentration profile emerging from the organ is lagged,
unimodal, asymmetric, and skewed to the right (Roberts et al., 1988
;
Krejcie et al., 1996
). This shape arises from dispersion of the
idealized impulse input in the organ due to laminar flow in the
vasculature, turbulence at branch points, the distribution of vascular
path lengths, and heterogeneity in the perfusion of the organ (Krejcie
et al., 1996
). More complex, stochastic "dispersion" models of the
organ are required to account for these processes, often with models
that need to be solved by computationally intensive inversion of the Laplace solution to the model (Roberts et al., 1988
).
It would be advantageous to find the middle ground between the
computational simplicity but lower fidelity of a single-compartment model, and the opposite is true for stochastic dispersion-based models.
The data suggest that a class of compartmental models known as "tank
in series" models (Beek and Muttzall, 1975
; Roberts et al., 1988
) may
provide a means of doing this, as they can represent increased
dispersion by increasing the number of compartments in series. However,
these models can be written as differential equations in the time
domain and, therefore, solved in the same manner as the traditional
single-compartment organ models used in many physiological models.
The effect of adding extra compartments in these "in series"
and "peripheral" configurations is not immediately obvious.
However, the simulations shown in Fig. 6
illustrate the general principles. Note that each configuration has a
characteristic effect on the venous concentrations emerging from the
model. First, the single-compartment model is characterized by an
exponential rise and fall of the venous concentrations when
exposed to a "square wave" arterial concentration input.
Importantly, the peak venous concentration always occurs at the point
at which the venous concentrations cross the arterial
concentrations
this is an embodiment of Zilversmit's rule originally
developed for the analysis of product-precursor relationships (Rescigno
and Segre, 1966
; Jones and Nicholas, 1984
). This behavior is
characteristic of systems in which there is only one compartment
directly between arterial and venous blood. Second, it is clear that
the principal advantage of the tank in series configuration is that it
is possible to introduce a delay in the time of the peak venous
concentration. The peak is also skewed to the right in a manner
consistent with real data (Krejcie et al., 1996
). Finally, note that
the peripheral configuration of membrane-limited models also must obey
Zilversmit's rule by nature of the arrangement of their compartments,
and the peak venous concentration will always occur when the venous
concentrations cross the arterial, regardless of the number of
additional peripheral compartments. Indeed, the principal effect of
adding additional peripheral compartments is to prolong the elution
process with only minor changes in the rate of uptake. The advantage of
the compilation models is that both the delayed peak characteristic of
dispersion and the prolonged elution characteristic of deep distribution can be accounted for to different extents as
determined by the data.
|
An overall conclusion from this analysis is that arteriovenous
concentration difference data across an organ can be used to differentiate between a variety of kinetic models. The redundancy of
compartmental models when fitted to systemic blood concentration data,
as pointed out by Wagner (1975)
, is not an issue in this quite
different experimental paradigm. Overall, compilation model A was
considered the optimal model (from those examined) for these data
because it was the best fit to the kinetic data and one of the best for
the kinetic-dynamic data without nonidentifiable parameters. This model
is of a novel form because it incorporates both dispersion (two
compartments in series) and deep distribution elements (a peripheral
compartment). We currently are investigating the hypothesis that this
relatively simple type of model may be of a sufficiently general form
that it is suitable for describing regional kinetics for a wide range
of organ and drug combinations.
| |
Footnotes |
|---|
Accepted for publication March 30, 1999.
Received for publication October 20, 1998.
1 This work was supported by research grants from the National Heart Foundation of Australia and the National Health and Medical Research Council of Australia.
2 Current address: Cardiac Technology, Royal North Shore Hospital, University of Sydney, St. Leonards, NSW 2065, Australia.
3 Current address: Department of Anesthesia and Pain Management, Royal North Shore Hospital, University of Sydney, St. Leonards, NSW 2065, Australia.
Send reprint requests to: Dr. Richard N. Upton, Department of Anaesthesia and Intensive Care, Royal Adelaide Hospital, University of Adelaide, Adelaide, Australia. E-mail: rupton{at}health.adelaide.edu.au
| |
Abbreviations |
|---|
CNS, central nervous system; LV, left ventricle; LV dP/dtmax, maximum positive rate of change of left ventricular pressure; AUC, area under the curve; MSC, model selection criteria; Qh, myocardial blood flow; Cart, arterial meperidine concentration; Ccs, coronary sinus meperidine concentration.
| |
The Equations of the Kinetic Models |
|---|
|
|
|---|
The volume of the compartment drained by coronary sinus blood (concentration given by Ccs) was designated as Vh. The volume of the additional compartments was designated V1 and V2, with concentrations C1 and C2, respectively. The apostrophe symbol (') after a variable indicates a derivative with time. The volumes of distribution are apparent volumes; thus, the concentrations in each volume will be numerically equivalent at steady state. The kinetic-dynamic models added the dynamic equation discussed in the text, with the effect linked to each concentration in the compartment (Ccs, C1, or C2) in turn. The optimal model from these alternatives was chosen as the final kinetic-dynamic model.
Model 1: Single Flow-Limited Compartment
Vh *
Ccs' = Qh *
(Cart
Ccs)
Model 2: Two Tanks in Series
V1 *
C1' = Qh *
(Cart
C1)
Vh *
Ccs' = Qh *
(C1
Ccs)
Model 3: Traditional Membrane-Limited Model
Vh *
Ccs' = Qh *
(Cart
Ccs) + PS1 *
(C1
Ccs)
V1 *
C1' = PS1 *
(Ccs
C1)
Model 4: Three Tanks in Series
V1 *
C1' = Qh * (Cart
C1)
V2 *
C2' = Qh *
(C1
C2)
Vh *
Ccs' = Qh *
(C2
Ccs)
Model 5: Compilation A
V1 *
C1' = Qh *
(Cart
C1)
Vh *
Ccs' = Qh *
(C1
Ccs) + PS2 *
(C2
Ccs)
V2 *
C2' = PS2 *
(Ccs
C2)
Model 6: Compilation B
V1 *
C1' = Qh *
(Cart
C1) + PS2 *
(C2
C1)
V2 *
C2' = PS2 *
(C1
C2)
Vh *
Ccs' = Qh *
(C1
Ccs)
Model 7: Double Membrane-Limited A
Vh *
Ccs' = Qh *
(Cart
Ccs) + PS1 *
(C1
Ccs)
V1 *
C1' = PS1 *
(Ccs
C1) + PS2 *
(C2
C1)
V2 *
C2' = PS2 *
(C1
C2)
Model 8: Triangular
V1 *
C1' = Qh *
Cart
(Qh
Q2) *
C1
Q2 *
C1
V2 *
C2'
=Q2 *
C1
Q2 *
C2
Vh *
Ccs' = (Qh
Q2) *
C1
Qh *
Ccs + Q2 *
C2
Model 9: Double Membrane-Limited B
Vh *
Ccs' = Qh *
(Cart
Ccs) + PS1 *
(C1
Ccs) + PS2
* (C2
Ccs)
V1 *
C1' = PS1 *
(Ccs
C1)
V2 *
C2' = PS2 *
(Ccs
C2)
| |
References |
|---|
|
|
|---|
Confidence Intervals and Statistical Guidelines.
British Medical Journal, London.
Physiological analysis and modeling.
J Pharm Exp Ther
284:
1048-1057
-acetylmethadol (LAAM).
J Pharmacol Exp Ther
215:
259-265
Application of mass balance principles.
Anesth Pharmacol Rev
2:
231-242.
Relationship to thiopental induced reductions in myocardial contractility.
J Pharm Sci
85:
863-867[Medline].This article has been cited by other articles:
![]() |
R. N. Upton, C. Grant, A. M. Martinez, and G. L. Ludbrook Recirculatory model of fentanyl disposition with the brain as the target organ Br. J. Anaesth., November 1, 2004; 93(5): 687 - 697. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Zheng, R. N. Upton, G. L. Ludbrook, and A. Martinez Acute Cardiovascular Effects of Magnesium and Their Relationship to Systemic and Myocardial Magnesium Concentrations after Short Infusion in Awake Sheep J. Pharmacol. Exp. Ther., June 1, 2001; 297(3): 1176 - 1183. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||